Next: COUNTING YEAST CELLS
Up: Lab Techniques for the
Previous: USING THE SPEC-20 FOR
- Mark a paper towel with labels for each sample you will be
analyzing. (We will use this to store temporarily some of the
pipetting tips.)
- Remove and spin-down a sample from each flask.
- Set a P-1000 micropipetter dial to 50 (0.5 ml). Attach a
blue tip.
- Homogenize the sample by first swirling by hand then use
the Vortex shaker set at Shake 3 for 15 seconds.
- Without touching the sides or edge of the flask extract 0.5
ml from the flask and expell it into the plastic
microcentrifuge tupes (with caps).
- Label the cuvettes, and put them balanced in the
microcentrifuge. If the number of cuvettes is odd; add one
with water. Spin according to attached directions. Remove
and set aside for later use.
- Put 2 ml of glucose reagent in each Spec-20 cuvette (blank, standard,
samples).
- The reagent is kept in the refrigerator along the NORTH
wall.
- Use the bottle with the black dot on the lid so we
can best monitor usage for re-ordering.
- Use the P-1000 micropipette with the dial set to 100 and a
blue pipette tip. Without touching the sides of the flask,
extract 1 ml twice and deposit each in the cuvette.
- Blank: add nothing more.
- Standard: add 0.02 ml of glucose standard.
- Find the dropper
bottle of the standard and squeeze 1 small drop into cap.
- Attach a plastic tip to the P-20 micropipetter. Set the
dial to 200. Extract 0.02 mL of glucose standard into the
tip.
- Deposit the tip contents near the surface of the reagent in
the standard's cuvette. Depress the liquid expelling plunger firmly.
Touch the tip to the inside of cuvette to remove any traces.
Remove the pipette tip and place it on the proper place on the
paper towel.
- Sample: Repeat the above procedure using a new tip for 0.02 ml of the
sample in its cuvette.
- Mix each cuvette using the test-tube Vortex set at Vortex-1
for about 5 seconds. Then let everything sit for 10 min. After
incubation, read absorbance on the Spec-20 using the attached
instructions. Readings must be done within 60 minutes. The
contents of the standard cuvette should be a dark
pink.
- If the sample glucose is too concentrated (absorbance greater than
0.7), dilute with distilled water using a pipette until the
absorbance is less than 0.7. Note the dilution factor on your
data sheet.
- Record absorbance, dilution factor, and glucose concentration.
Glucose concentration (mg/L) is: (Sample absorbance)/(Standard
absorbance) x 1000.
Next: COUNTING YEAST CELLS
Up: Lab Techniques for the
Previous: USING THE SPEC-20 FOR
James Powell
2000-07-31